Digital Temperature
Measurement
|
|
The Series 460 measures and displays accurate temperatures reliably and repeatedly. Sensor input types are selectable from a smooth, environmentally protected front panel keypad. Choose either Model 460, a 4-digit resolution meter, or Model 461, a higher resolution 5-digit instrument. With either model, you can add a variety of input options to tailor the instrument precisely to your needs. For unmatched repeatability and accuracy, the Series 460 uses a new wide-range, integrating voltage-to-frequency conversion technique. An integral microprocessor generates periodic digital auto-zero readings for baseline verification. Because temperature
sensors emit very low signal levels, RFI and EMI often present the most
serious obstacle to accurate readings. The Series 460 common mode rejection
ratio exceeds 120dB, ensuring rejection of unwanted common mode voltages
from ground loops that would otherwise make measurements unreliable.
Outputs are also shielded, preserving data integrity all the way to
your readout or printout. |
|
Range Accuracy and Inputs |
|||
| Range Designator on Display |
Sensor Type | Range | Accuracy (all ±) |
| Thermocouple | |||
| J T/C | Iron-Constantan | -336 to 2193°F -204 to 1200°C 68 to 1473 K |
0.5°F 0.3°C 0.3 K |
| K T/C | Chromel®*-Alumel®* | -229 to 2503°F -184 to 1373°C 88 to 1646 K |
0.6°F 0.3°C 0.3 K |
| T T/C | Copper-Constantan | -373 to 753°F -225 to 400°C 48 to 674 K |
0.4°F 0.2°C 0.2 K |
| E T/C | Chromel®*-Constantan | -358 to 1835°F -217 to 1001°C 56 to 1275 K |
0.6°F 0.3°C 0.3 K |
| R T/C | Platinum-Platinum 13% Rhodium |
32 to 3219°F 0 to 1770°C 273 to 2043K |
1.2°F (>79.0°F) 0.7°C (>26.0°C) 0.7 K (>299.2 K) |
| S T/C |
Platinum-Platinum 10% Rhodium |
32 to 3216°F 0 to 1769°C 273 to 2042 K |
1.2°F (>79.0°F) 0.7°C (>26.0°C) 0.7 K (>299.0 K) |
| B T/C | Platinum 30% Rhodium Platinum 6% Rhodium |
899 to 3309°F 481 to 1820°C 755 to 2094K |
2.0°F 1.1°C 1.1 K |
| N T/C | Nicrosil-Nisil | 32 to 2301°F 0 to 1260°C 273 to 1534 K |
0.5°F 0.3°C 0.3 K |
| C T/C | Tungsten 5% Rhenium Tungsten 26% Rhenium |
-10 to 4200°F -23 to 2315°C 249 to 2589 K |
1.1°F 0.6°C 0.6 K |
| G T/C | Tungsten-Tungsten 26% Rhenium |
295 to 4206°F 146 to 2318°C 419 to 2592 K |
1.1°F 0.6°C 0.6 K |
| D T/C | Tungsten 3% Rhodium Tungsten 25% Rhodium |
-8.4 to 4203°F -22.4 to 2317°C 250 to 2590 K |
1.1°F 0.6°C 0.6 K |
| CGI T/C | Chromel®*-Gold .07% Atomic Iron |
-460 to 44°F -273 to 7°C 0 to 280 K |
0.4°F 0.2°C 0.2 K |
| FeCon | Iron Constantan DIN |
-330 to 1654°F -210 to 901°C 71 to 1174 K |
0.9°F 0.5°C 0.5 K |
| CuCon | Copper-Constantan DIN |
-331 to 1114°F -201 to 601°C 71 to 874 K |
0.9°F 0.5°C 0.5 K |
| *CHROMEL® and ALUMEL® are registered trademarks of the Hoskins Manufacturing Company. | |||
| RTD | |||
| PT1000 385 | Platinum-100 3- or 4-wire a=0.00385/DIN 43670 curve |
-332 to 1571°F -202 to 855°C 70 to 1128 K |
0.2°F 0.1°C 0.1 K |
| PT100 392 | Platinum-100 3 or 4-wire a=0.00392 |
-410 to 1572°F -240 to 855°C 32 to 1128 K |
0.2°F 0.1°C 0.1 K |
| PT200 392 | Platinum-200 3 or 4-wire a=0.00392 |
-330 to 1566°F -201 to 852°C 72 to 1125 K |
0.2°F 0.1°C 0.1 K |
| PT1000 | Platinum-100 3- or 4-wire a=0.00375/HYCAL Ultra 7 curve |
-330 to 1025°F -204 to 1200°C 68 to 1473 K |
0.2°F 0.1°C 0.1 K |
| Cu10 | Copper-10 3 or 4-wire per Minco table 16-9 |
-337 to 507°F -205 to 264°C 68 to 537 K |
0.8°F 0.4°C 0.4 K |
| Ni120 | Nickel-120 3 or 4-wire per Minco table 7-120 |
-112 to 608°F -80 to 320°C 192 to 593 K |
0.4°F 0.2°C 0.2 K |
| Thermistor | |||
| YSI 400 | Thermistor Series YSI 400 2-wire |
-42.4 to 221.7°F -41.3 to 185.4°C 231.9 to 458.6 K |
0.4°F 0.2°C 0.2 K |
| YSI 700 | Thermistor Series YSI 700 3-wire |
31.0 to 213.0°F -0.6 to 100.6°C 272.6 to 373.8 K |
0.4°F 0.2°C 0.2 K |
|
Series 460 Options |
| Tailor
the instruments with options to extend your measurement capabilities.
Five option slots are available in each meter. The following options can
be specified now, or added at any time in the future.
Multiple Inputs: Monitor as many as 24 inputs. The first multiple input card switches six channels of thermocouples or three channels of RTD or thermistor units. Maximum of 4 cards per instrument. Order Option: -20 (6 inputs of thermocouple); -21 (3 inputs of RTD/thermistor). Math Expressions: Provides display of maximum, minimum, rate of change or timed average value of measurements. Front panel keys allow you to view the math value or switch back to view the actual measured value. Order Option: -04 Alarms: Each option card provides two alarms with separate limits and relays. Up to two cards can be specified. Any combination of high, low or ± values can be used to trigger the alarm, from either actual or math processed input. Trigger delay and a settable dead-band are included as standard. Order Option: -03 (order two for 4 alarms) Analog Output: For output to recorders or other analog input instruments, you can choose from two fully isolated, linearized analog outputs. Both output types, 4-20mA or 0-10VDC can be fully scaled. Output can be assigned to direct measurements or to math values. Order Option: -06 (4-20mA); -05 (0-10VDC) Digital Output: Serial Output RS232C or 20mA interface sends display information in serial ASCII format. Includes user selectable transmission rates from 300 to 9600 baud, line feed, and carriage return. Order Option: -07 IEEE-488 Output. For integration with IEEE-488 Bus based measurement or analytical systems. Main indicator and other options can be programmed via the IEEE-488 Bus. (Requires two option slots.) Order Option: -08 Differential measurement. This card provides an additional input channel and the capability to display the difference (sT) between the two channels. This option card, -22, can be specified with thermocouple and voltage input models. The -23 option is specifically for use with RTD or thermistor configurations. This function can
be combined with other options: maximum, minimum, and average deltas
when used with math capability, scaled analog output of differentials
when used with analog output, and alarm points based on differential
limits (Cannot be used in same indicator with -20 and -21 options. Not
for use with current loop or strain-gage versions.) Order Option
-22 for DELTA and 2 thermocouple or voltage inputs, -23 for DELTA and
2 RTD or thermistor inputs. |